+606-601 6457


DIROSYS SOLUTIONS
(M) SDN BHD


No. 197, Jalan S2 E9,
Garden Homes Seremban 2,
70300 Seremban,
Negeri Sembilan, Malaysia.

+606-601 6457
+606-601 6451

Contact :
MR. CH LAM
+6012-663 8622
+6017-369 7963

Automatic Dialyzers Reprocessing System

Automatic Dialyzers Reprocessing System

Automatic Dialyzers Reprocessing System

Automatic dialyzer Reprocessor
  • Fully automatic reprocessing and disinfecting cycles
  • Built-in anti-splash shield
  • Chemical detection sensor
  • Dialyzerdis infectant fill sensor
  • Membrane pressure leak test
  • Chemical concentration sensor
Technical Spec
  • Power: 100-130 VAC220-240VAC
  • Consumption: 150W
  • Dimension: H420 x W380x D420mm
Features and Benefits
 
Reliable and Safe
  • Fully automatic reprocessing and disinfecting cycles
  • Built-in effluent basin
  • Membrane pressure leak test
  • Low water inlet pressure alarm
  • Chemical detection sensor
  • Dialyzer disinfectant fill sensor
  • Chemical concentration sensor
User-Friendly
  • Easy to operate LCD touch screen
  • Select able or presets of dialyzer cell volume selections
  • Total cell volume measurement and display
  • Reverse ultra-filtration cleaning
  • Selectable programs for all types of dialyzer Innovative Design
  • Economical consumption of disinfectant
  • Accepts most common disinfectants
 
Installation
Environmental Limits
Ambient temperature: 18-35º Celsius 
Relative humidity: up to 80%

Water requirements
Water quality: must comply to AAMI RD62
Water pressure: 20-50 PSI
Water flow: min. 1.5 L/min
Temperature: 20-35º Celsius
Consumption: ≈18 L/cycle

Drain requirements
Flow: min. 3.5L/min
Height: min. 200mm below TEC11

Technical Specifications
Dimensions: Height: 420mm
Width: 380mm
Depth: 420mm
Weight: ~30kg
Power: 100 –130 VAC, 60Hz220 –240 VAC, 50Hz
Consumption: 150W

Performance
Dialyzer Testing
Pressure Leak Test:
Pressure tested to 300mmHg ±10mmHg
Acceptable leakage rate is 10%
Volume Testing:
Range 10ml to 500ml
Accuracy ±5%

Reprocessing Time
Approximately 6 -12minutes per dialyzer

Disinfectants
Company recommended disinfectants

Alarms
Audio / Visual indication of:
• Low water inlet pressure
• Low disinfectant
• Disinfectant concentration (high & low)
• Leak test failure
• Total cell volume high & low alarm preset
• Cycle complete message
• Volume measurement error
 
 

Download Files

Please leave your enquiry here, we will reply as soon as possible.
Name*  
Company Name  
Product Interested  
Quantity  
Email*  
Contact No.*  
Attachment  
*only support gif, jpeg, jpg, png, pdf
Messages*  

Switch to Mobile Version